Assembly Replacement Sequence
The following steps show the sequence to replace an assembly in the analyzer.
1. Identify the faulty group by following the procedures in this chapter. Follow up with the
appropriate troubleshooting chapter that identifies the faulty assembly.
2. Order a replacement assembly. Refer to
3. Replace the faulty assembly and determine what adjustments are necessary. Refer to
Chapter 14 , "Assembly Replacement and Post-Repair Procedures."
4. Perform the necessary adjustments. Refer to
Constants."
5. Perform the necessary performance tests. Refer to
Chapter 4
Chapter 13 , "Replaceable Parts."
Chapter 3 , "Adjustments and Correction
Chapter 2 , "Performance Tests."
Start Troubleshooting Here
Assembly Replacement Sequence
4-3