Assembly Replacement and Post-Repair Procedures

A15 Preregulator

A15 Preregulator
Tools Required
• T-10 TORX screwdriver
• T-15 TORX screwdriver
• ESD (electrostatic discharge) grounding wrist strap
Removal
1. Remove the rear panel (refer to
2. Remove the two remaining screws from the top of the rear frame.
3. Disconnect the wire bundle (A15W1) from A8J2 and A17J3.
4. Remove the preregulator (A15) from the frame.
Replacement
1. Reverse the order of the removal procedure.
NOTE
When reinstalling the preregulator (A15), make sure the three grommets
(item 1) on A15W1 are seated in the two slots (item 2) on the back side of the
preregulator and the slot (item 3) in the card cage wall.
After reinstalling the preregulator (A15), be sure to set the line voltage
selector to the appropriate setting, 115 V or 230 V.
14-30
"Rear Panel Assembly" on page
14-14).