8. Receiver Phase Frequency Response (Ratio)
Perform this test to determine the phase tracking frequency response for each pair of
inputs (A/R, B/R and A/B) in the swept sweep mode.
Analyzer warmup time: 30 minutes.
Specifications
Frequency Range
300 kHz–3 GHz
c
3 GHz–6 GHz
a. Deviation from linear phase. IF BW ≤ 300 Hz and 3 second
sweeptime for A/B measurements.
b. At 25 °C ± 5 °C, with −10 dBm into all receiver inputs.
c. Only for analyzers with Option 006.
Required Equipment
Description
2-way power splitter (2), 50Ω
RF cable set, 50Ω Type-N
RF cable, 50Ω Type-N
50 Ω Termination
Procedure
A/R Phase Frequency Response
1. Connect the equipment as shown in
RF cable set connected to port A.
Chapter 2
Phase Frequency Response
± 3°
± 10°
HP/Agilent Part or Model Number
11667A, Option 001
11851B
8120-4781
from 85032B Type-N calibration kit
Figure 2-12
8. Receiver Phase Frequency Response (Ratio)
a, b
with the second type-N cable from the
Performance Tests
2-27