SOURCE POWER TURNED OFF, RESET UNDER POWER MENU
Information
You have exceeded the maximum power level at one of the inputs and power has been
Message
automatically reduced. The annotation P⇓ indicates that power trip has been
activated. When this occurs, reset the power and then press
SOURCE PWR on OFF
numbers 57, 58, and 59.
SWEEP MODE CHANGED TO CW TIME SWEEP
Error Number 187
If you select external source auto or manual instrument mode and you do not also
select CW mode, the analyzer is automatically switched to CW.
TEST ABORTED
Error Number 113
You have prematurely stopped a service test.
TROUBLE!CHECK SETUP AND START OVER
Service Error
Your equipment setup for the adjustment procedure in progress is not correct. Check
Number 115
the setup diagram and instructions in
(Test 48)" on page 3-10
WRONG DISK FORMAT, INITIALIZE DISK
Error Number 77
You have attempted to store, load, or read file titles, but your disk format does not
conform to the Logical Interchange Format (LIF). You must initialize the disk before
reading or writing to it.
Chapter 10
, to switch on the power. This message follows error
"Source Pretune Correction Constants
. Start the procedure again.
Service Key Menus and Error Messages
Error Messages
Power
10-49