Performance Tests
9. Receiver Input Crosstalk
33.Press
Start
300
34.Press
Meas
A/R
.
SINGLE
35.At the end of the sweep, press
.
B/R
36.Press
AVERAGING FACTOR
Avg
TRIGGER MENU
37.When the analyzer finishes the number of sweeps, write the marker value, which
appears on the analyzer display, on the performance test record.
A into B Crosstalk from 1 GHz to 3 GHz
38.Press
Start
1
39.Press
Meas
A/R
.
SINGLE
40.At the end of the sweep, press
.
B/R
41.Press
AVERAGING FACTOR
Avg
TRIGGER MENU
42.When the analyzer finishes the number of sweeps, write the marker value, which
appears on the analyzer display, on the performance test record.
(Only for Option 006 Analyzers) A into B Crosstalk from 3 GHz to 4.5 GHz
43.Press
Start
3
44.Press
Meas
A/R
.
SINGLE
45.At the end of the sweep, press
.
B/R
46.Press
AVERAGING FACTOR
Avg
TRIGGER MENU
47.When the analyzer finishes the number of sweeps, write the marker value, which
appears on the analyzer display, on the performance test record.
(Only for Option 006 Analyzers) A into B Crosstalk from 4.5 GHz to 6 GHz
48.Press
Start
4.5
49.Press
Meas
A/R
.
SINGLE
50.At the end of the sweep, press
.
B/R
2-36
k/m
Stop
1
G/n
Avg
AVERAGING OFF
Display
5
NUMBER of GROUPS
G/n
Stop
3
G/n
Avg
AVERAGING OFF
Display
5
NUMBER of GROUPS
G/n
Stop
4.5
G/n
Avg
AVERAGING OFF
Display
5
NUMBER of GROUPS
G/n
Stop
6
G/n
Avg
AVERAGING OFF
Display
.
Sweep Setup
DATA → MEMORY DATA/MEM
AVERAGING ON
x1
.
5
x1
.
Sweep Setup
DATA → MEMORY DATA/MEM
AVERAGING ON
x1
.
5
x1
.
Sweep Setup
DATA → MEMORY DATA/MEM
AVERAGING ON
x1
.
5
x1
.
Sweep Setup
DATA → MEMORY DATA/MEM
TRIGGER MENU
Meas
Sweep Setup
TRIGGER MENU
Meas
Sweep Setup
TRIGGER MENU
Meas
Sweep Setup
TRIGGER MENU
Meas
Chapter 2