Performance Tests

14. Receiver Phase Compression

14. Receiver Phase Compression
Perform this test to verify the compression/expansion magnitude levels of the analyzer's
receiver samplers.
Analyzer warmup time: 30 minutes
Specifications
Frequency Range
300 kHz–3 GHz
b
3 GHz–6 GHz
300 kHz–3 GHz
b
3 GHz–6 GHz
a. With 10 Hz IF bandwidth.
b. Only for analyzers with Option 006.
Required Equipment
Description
Power Meter
Power Sensor
Attenuator, 3 dB
Attenuator, 10 dB
Attenuator, 20 dB
Cable (2), 50 Ω type-N, 24-inch
2-Way power splitter (2), 50 Ω
Additional Equipment for Option 006 Analyzers
Power Sensor
Channel A Phase Compression
1. Zero and calibrate the power meter.
2. Connect the equipment as shown in
2-62
Receiver
Compression
Channel
≤ 8.0 °
A
≤ 8.1 °
A
≤ 8.0 °
B
≤ 8.1 °
B
HP/Agilent Part or Model Number
437B/438A
8482A
8491A Option 003
8491A Option 010
8491A Option 020
8120-4781
11667A Option 001
8481A
Figure
a
2-29.