Figure 11-159P-Series Sensor Error In Peak-To-Average Measurements For A Two-Tone Input - Agilent Technologies N1911A User Manual

P-series power meters
Hide thumbs Also See for N1911A:
Table of Contents

Advertisement

11
Specifications and Characteristics
Characteristic Peak Flatness
The peak flatness is the flatness of a peak- to- average ratio measurement
for various tone- separations for an equal magnitude two- tone RF input.
Figure 11- 159
measurements as the tone separation is varied. The measurements were
performed at –10 dBm with power sensors with 1.5 m cable lengths.
0.5
0.0
-0.5
-1.0
-1.5
-2.0
-2.5
-3.0
-3.5
0

Figure 11-159P-Series sensor error in peak-to-average measurements for a two-tone input

268
refers to the relative error in peak- to- average ratio
OFF
( < 500MHz)
LOW
5
10
Input tone separation frequency (MHz)
MEDIUM
15
20
N1911A/1912A P-Series Power Meters User's Guide
HIGH
OFF
( > 500MHz)
25
30