Adjustments and Correction Constants
IF Amplifier Correction Constants (Test 51)
IF Amplifier Correction Constants (Test 51)
This adjustment routine measures the gain of the IF amplifiers (A and B only) located on
the A10 digital IF, to determine the correction constants for absolute amplitude accuracy.
Analyzer warmup time: 30 minutes.
Required Equipment and Tools
Description
Power splitter
RF cable (2)
Attenuator, 20 dB, type-N
Antistatic wrist strap
Antistatic wrist strap cord
Static-control table mat and earth ground wire
1. Connect the equipment as shown in
Figure 3-5
Setup for IF Amplifier Correction Constants
2. Press
Preset
System
.
CONTINUE
3. Connect the equipment as shown in
3-16
HP/Agilent Part or Model Number
11667A Option 001
11500B
8491A, Option 020
9300-1367
9300-0980
9300-0797
Figure
3-5, using the analyzer's A input.
SERVICE MENU TESTS
Figure
3-5, using the analyzer's B input.
EXECUTE TEST YES
51
x1
Chapter 3