Eeprom Backup Disk Procedure - Agilent Technologies 8753ES Option 011 Service Manual

Network analyzer
Hide thumbs Also See for 8753ES Option 011:
Table of Contents

Advertisement

EEPROM Backup Disk Procedure

The correction constants, which are unique to your instrument, are stored in EEPROM on
the A9 controller assembly. By creating an EEPROM backup disk, you will have a copy of
all the correction constant data should you need to replace or repair the A9 assembly.
Required Equipment and Tools
Description
3.5-inch Floppy Disk
Antistatic Wrist Strap
Antistatic Wrist Strap Cord
Static-control Table Mat and Ground Wire
Procedure
1. Insert a 3.5-inch disk into the analyzer disk drive.
2. If the disk is not formatted, follow these steps:
a. Press
Save/Recall
b. Select the format type:
• To format a LIF disk, select
• To format a DOS disk, select
c. Press
FORMAT INT DISK
3. Press
SERVICE MENU
System
Save/Recall
SELECT DISK
The analyzer creates a default file "FILE0". The file name appears in the
NOTE
upper-left corner of the display. The file type "ISTATE(E)" indicates that the
file is an instrument-state with EEPROM backup.
4. Press
FILE UTILITIES
and the
SELECT LETTER
"FILE0" TO "N12345" where 12345 represents the last 5 digits of the instrument's
serial number. (The first character in the file name must be a letter.) When you are
finished renaming the file, press
5. Write the following information on the disk label:
• analyzer serial number
• today's date
• "EEPROM Backup Disk"
Chapter 3
HP/Agilent Part or Model Number
One formatted 1.44 MB disk
9300-1367
9300-0980
9300-0797
FILE UTILITIES
FORMAT DISK
FORMAT:LIF
FORMAT:DOS
and answer
SERVICE MODES
INTERNAL DISK
RENAME FILE
ERASE TITLE
softkey (or an external keyboard) to rename the file
.
DONE
Adjustments and Correction Constants
EEPROM Backup Disk Procedure
.
.
.
at the query.
YES
MORE
RETURN
SAVE STATE
. Use the front panel knob
STORE EEPR ON
.
3-33