Adjustments and Correction Constants

ADC Offset Correction Constants (Test 52)

ADC Offset Correction Constants (Test 52)
Analyzer warmup time: 30 minutes.
These correction constants improve the dynamic accuracy by shifting small signals to the
most linear part of the ADC quantizing curve.
1. Press
Preset
System
.
YES
NOTE
This routine takes about three minutes.
2. Observe the analyzer for the results of the adjustment routine:
• If the analyzer displays ADC Ofs Cor DONE, you have completed this procedure.
• If the analyzer displays ADC Ofs Cor FAIL, refer to
Troubleshooting."
3-18
SERVICE MENU
TESTS
52
x1
EXECUTE TEST
Chapter 6 , "Digital Control
Chapter 3