Source Pretune Default Correction Constants (Test 45) - Agilent Technologies 8753ES Option 011 Service Manual

Network analyzer
Hide thumbs Also See for 8753ES Option 011:
Table of Contents

Advertisement

Adjustments and Correction Constants

Source Pretune Default Correction Constants (Test 45)

Source Pretune Default Correction Constants (Test 45)
Analyzer warmup time: 30 minutes.
This adjustment writes default correction constants for rudimentary phase lock pretuning
accuracy.
1. Press
Preset
System
.
YES
2. Observe the analyzer for the results of this adjustment routine:
• If the analyzer displays Pretune Def DONE, you have completed this procedure.
• If the analyzer displays FAIL, refer to
3-8
SERVICE MENU
TESTS
Chapter 7 , "Source Troubleshooting."
45
x1
EXECUTE TEST