Agilent Technologies 8753ES Option 011 Service Manual page 68

Network analyzer
Hide thumbs Also See for 8753ES Option 011:
Table of Contents

Advertisement

Performance Tests
12. Receiver Magnitude Dynamic Accuracy
Figure 2-22
Power Meter Calibration
8. Set the 8496A to 20 dB.
9. Set the following analyzer parameters:
Preset
Sweep Setup
NUMBER of POINTS
Power
10
IF BW
Avg
10.Set up the analyzer for power meter calibration:
a. Select the analyzer as the system controller:
Local
SYSTEM CONTROLLER
b. Set the power meter's address:
SET ADDRESSES
ADDRESS: P MTR/GPIB
c. Select the appropriate power meter by pressing
model number is displayed. (Use the 438A/437 selection if the power meter is an
HP/Agilent E4419B or E4418B.)
d. Select the calibration kit and enter the power sensor calibration data.
CAL KIT
Cal
PWRMTR CAL
Cal
the power sensor calibration data for 30 MHz)
2-52
CW FREQ
51
x1
x1
10
x1
13
SELECT CAL KIT
LOSS/SENSOR LISTS
30
M/µ
x1
POWER MTR
7mm 85031
CAL FACTOR SENSOR A
DONE
until the correct
(enter
Chapter 2