Adjustments and Correction Constants

Correction Constants Retrieval Procedure

Correction Constants Retrieval Procedure
Required Equipment and Tools
Description
EEPROM backup disk
Antistatic wrist strap
Antistatic wrist strap cord
Static-control table mat and earth ground wire
By using the current EEPROM backup disk, you can download the correction constants
data into the instrument EEPROMs.
1. Insert the "EEPROM Backup Disk" into the analyzer's disk drive.
2. Make sure the A9 switch is in the Alter position.
3. Press
Save/Recall
highlight the file "N12345" where N12345 represents the file name of the EEPROM
data for the analyzer. On the factory shipped EEPROM backup disk, the file name is
"FILE1".
4. Press
RETURN
instrument EEPROMs.
5. Press
and verify that good data was transferred to EEPROM by performing a
Preset
simple measurement.
6. Move the A9 switch back to the Normal position when you are done working with the
instrument.
3-34
SELECT DISK
INTERNAL DISK
to download the correction constants data into the
RECALL STATE
HP/Agilent Part Number
9300-1367
9300-0980
9300-0797
. Use the front panel knob to