If The Green Led Of The A15 Is Not On Steadily - Agilent Technologies 8753ES Option 011 Service Manual

Network analyzer
Hide thumbs Also See for 8753ES Option 011:
Table of Contents

Advertisement

If the Green LED of the A15 Is Not ON Steadily

If the green LED is not on steadily, the line voltage is not enough to power the analyzer.
Check the Line Voltage, Selector Switch, and Fuse
Check the main power line cord, line fuse, line selector switch setting, and actual line
voltage to see that they are all correct.
a small flat-blade screwdriver to pry out the fuse holder.
the line voltage selector switch. Use a small flat-blade screwdriver to select the correct
switch position.
NOTE
Refer to
description and part number.
If the A15 green LED is still not on steadily, replace A15.
Figure 5-4
Removing the Line Fuse
Chapter 5
Figure 5-4
"Hardware, Preregulator" on page 13-33
Power Supply Troubleshooting
If the Green LED of the A15 Is Not ON Steadily
shows how to remove the line fuse, using
Figure 5-2
shows the location of
for the correct line fuse
5-7