Adjustments and Correction Constants
Sampler Magnitude and Phase Correction Constants (Test 53)
9. Update the offsets for each sampler:
A-Channel Sampler
a. Access the first address by pressing
PEEK/POKE ADDRESS
b. Enter the new value for the A-sampler at the accessed address by pressing
[new value for A]
c. Press
Preset
B-Channel Sampler
a. Access the first address by pressing
PEEK/POKE ADDRESS
b. Enter the new value for the B-sampler at the accessed address by pressing
[new value for B]
c. Press
Preset
R-Channel Sampler
a. Access the first address by pressing
PEEK/POKE ADDRESS
b. Enter the new value for the R-sampler at the accessed address by pressing
[new value for R]
c. Press
Preset
10.Repeat this adjustment procedure.
11.If the analyzer continues to fail this adjustment routine, refer to
Troubleshooting."