Agilent Technologies 8753ES Option 011 Service Manual page 50

Network analyzer
Hide thumbs Also See for 8753ES Option 011:
Table of Contents

Advertisement

Performance Tests
9. Receiver Input Crosstalk
14.Press
B/R
Meas
.
x1
15.When the analyzer finishes the number of sweeps, write the marker value, which
appears on the analyzer display, on the performance test record.
R into B Crosstalk from 1 GHz to 3 GHz
16.Press
Start
1
.
5
x1
17.When the analyzer finishes the number of sweeps, write the marker value, which
appears on the analyzer display, on the performance test record.
(Only for Option 006 Analyzers) R into B Crosstalk from 3 GHz to 4.5 GHz
18.Press
Start
3
.
5
x1
19.When the analyzer finishes the number of sweeps, write the marker value, which
appears on the analyzer display, on the performance test record.
(Only for Option 006 Analyzers) R into B Crosstalk from 4.5 GHz to 6 GHz
20.Press
Start
4.5
.
5
x1
21.When the analyzer finishes the number of sweeps, write the marker value, which
appears on the analyzer display, on the performance test record.
B into A Crosstalk from 300 kHz to 1 GHz
22.Connect the equipment as shown in
Figure 2-15
B into A Receiver Input Crosstalk Test Setup
23.Press
Start
300
2-34
TRIGGER MENU
Sweep Setup
G/n
Stop
3
G/n
G/n
Stop
4.5
G/n
G/n
Stop
6
G/n
Figure
k/m
Stop
1
G/n
NUMBER of GROUPS
TRIGGER MENU
NUMBER of GROUPS
TRIGGER MENU
TRIGGER MENU
2-15.
.
5
NUMBER of GROUPS
NUMBER of GROUPS
Chapter 2