System Verification Tests These tests apply mainly to system-level, error-corrected
verification and troubleshooting. Tests 27 to 31 are associated with the system verification
procedure, documented in
examining the calibration coefficient arrays (error terms) resulting from a measurement
calibration; refer to
Chapter 11 , "Error Terms,"
Table 10-7
System Verification Tests
Test
Test Name
Number
27
Sys Ver Init
28
Ver Dev 1
29
Ver Dev 2
30
Ver Dev 3
31
Ver Dev 4
32–43
Cal Coef 1–12
Chapter 10
Chapter 2 , "Performance Tests."
Recalls the initialization state for system verification from an 8753ES
verification disk, in preparation for a measurement calibration. It must be done
before service internal tests 28, 29, 30, or 31 are performed.
Recalls verification limits from disk for verification device #1 in all applicable
measurements. It performs pass/fail limit testing of the current measurement.
Same as 28 above for device #2.
Same as 28 above for device #3.
Same as 28 above for device #4.
Copies error term data from a measurement calibration array to display
memory. A measurement calibration must be complete and active. The
definition of calibration arrays depends on the current calibration type. After
execution, the memory is automatically displayed. Refer to
"Error Terms,"
for details.
Service Key Menus and Error Messages
Tests 32 to 43 facilitate
for details.
Description
Service Key Menus
Chapter 11 ,