Accessories Troubleshooting
Accessories Troubleshooting
Accessories Troubleshooting
Use this procedure only if you have read
the procedures in the order given, unless instructed otherwise.
Measurement failures can be divided into two categories:
• Failures which don't affect the normal functioning of the analyzer but render incorrect
measurement data.
• Failures which impede the normal functioning of the analyzer or prohibit the use of a
feature.
This chapter addresses the first category of failures which are usually caused by the
following:
• operator errors
• faulty calibration devices or connectors
• bad cables or adapters
• improper calibration techniques
• RF cabling problems within the test set.
These failures are checked using the following procedures:
"Inspect the Accessories" on page 9-4
"Inspect the Error Terms" on page 9-5
"Test Set Troubleshooting" on page 9-8