Source Correction Routine
20.Press
DONE
RETURN
21.Press
RETURN
22.Connect the equipment as shown in
requested by the prompt.
Figure 3-4
Setup C for RF Output Correction Constants
23.Press
CONTINUE
24.Observe the analyzer display for the results of the adjustment routine:
• If the analyzer shows SOURCE Cor DONE, press
procedure.
• If the analyzer fails this routine, refer to
Chapter 3
PWR LOSS ON
TESTS
EXECUTE TEST
47
x1
Figure
.
Adjustments and Correction Constants
RF Output Power Correction Constants (Test 47)
to turn power loss ON.
.
YES
3-4, using splitter #2 and the power sensor
and you have completed this
Preset
Chapter 7 , "Source Troubleshooting."
3-15