Source Default Correction Constants (Test 44)

Analyzer warmup time: 30 minutes.
This internal adjustment routine writes default correction constants for the source power
accuracy.
1. Press
Preset
System
.
YES
2. Observe the analyzer for the results of the adjustment routine:
• If the analyzer displays *Source Def DONE, you have completed this procedure.
• If the analyzer displays *Source Def FAIL, refer to
Troubleshooting."
Chapter 3
SERVICE MENU
TESTS
Adjustments and Correction Constants
Source Default Correction Constants (Test 44)
44
x1
EXECUTE TEST
Chapter 7 , "Source
3-7