Adjustments and Correction Constants
Sampler Magnitude and Phase Correction Constants (Test 53)
Figure 3-7
Input A Sampler Correction Setup
16.Press
CONTINUE
17.Follow the analyzer prompt to move the cable from input A to input B, as shown in
Figure
3-8.
Figure 3-8
Input B Sampler Correction Setup
18.Press
CONTINUE
19.When the analyzer completes the adjustment, observe the display:
• If the analyzer shows DONE, this procedure is complete.
• If the analyzer shows FAIL, it is necessary to adjust the sampler offset values, which
are stored in EEPROM. Each sampler will require its own offset which is determined
by performing the following procedure for each sampler.
3-22
. The analyzer starts the second part of the automatic adjustment.
. The analyzer starts the third part of the automatic adjustment.
Chapter 3