Agilent Technologies 8753ES Option 011 Service Manual page 51

Network analyzer
Hide thumbs Also See for 8753ES Option 011:
Table of Contents

Advertisement

24.Press
A/B
Meas
.
x1
25.When the analyzer finishes the number of sweeps, write the marker value, which
appears on the analyzer display, on the performance test record.
B into A Crosstalk from 1 GHz to 3 GHz
26.Press
Start
1
.
5
x1
27.When the analyzer finishes the number of sweeps, write the marker value, which
appears on the analyzer display, on the performance test record.
(Only for Option 006 Analyzers) B into A Crosstalk from 3 GHz to 4.5 GHz
28.Press
Start
3
.
5
x1
29.When the analyzer finishes the number of sweeps, write the marker value, which
appears on the analyzer display, on the performance test record.
(Only for Option 006 Analyzers) B into A Crosstalk from
4.5 GHz to 6 GHz
30.Press
Start
4.5
.
5
x1
31.When the analyzer finishes the number of sweeps, write the marker value, which
appears on the analyzer display, on the performance test record.
A into B Crosstalk from 300 kHz to 1 GHz
32.Reconfigure the equipment as shown in
Figure 2-16
A into B Receiver Input Crosstalk Test Setup
Chapter 2
TRIGGER MENU
Sweep Setup
G/n
Stop
3
G/n
G/n
Stop
4.5
G/n
G/n
Stop
6
G/n
NUMBER of GROUPS
TRIGGER MENU
TRIGGER MENU NUMBER of GROUPS
TRIGGER MENU
Figure
2-16.
Performance Tests
9. Receiver Input Crosstalk
5
NUMBER of GROUPS
NUMBER of GROUPS
2-35