Analog Bus Codes
Measures and displays the analog input. The preset state input to
ANAI[D]
the analog bus is the rear panel AUX IN. The other 30 nodes may
be selected with
Outputs the counter's frequency data.
OUTPCNTR
Reads any prompt message sent to the error queue by a service
OUTPERRO
routine.
Outputs the integer status of the test most recently executed.
OUTPTESS
Status codes are those listed under "TST?".
Executes the power-on self test (internal test 1) and outputs an
TST?
integer test status. Status codes are as follows:
0 =pass
1 =fail
2 =in progress
3 =not available
4 =not done
5 =done