25.Press
MARKER 1
Marker
marker 2 to 300 kHz and 2 MHz respectively.
26.Use the analyzer front panel knob to move marker 2 to the peak value between 300 kHz
and 2 MHz. Write the marker 2 reading on the performance test record under "A Return
Loss (B/R)," 300 kHz–2 MHz.
27.Press
MARKER 1
Marker
marker 2 to 2 MHz and 1.3 GHz respectively.
28.Use the analyzer front panel knob to move marker 2 to the peak value between 2 MHz
and 1.3 GHz. Write the marker 2 reading on the performance test record under "A
Return Loss (B/R)," 2 MHz–1.3 GHz.
29.Press
MARKER 1
to 1.3 GHz and 3 GHz respectively.
30.Use the analyzer front panel knob to move marker 2 to the peak value between 1.3 GHz
and 3 GHz. Write the marker 2 reading on the performance test record under "A Return
Loss (B/R)," 1.3–3 GHz.
31.For Option 006:
a. Press
MARKER 1
to 3 GHz and 6 GHz respectively.
b. Use the analyzer front panel knob to move marker 2 to the peak value between
3 GHz and 6 GHz. Write the marker 2 reading on the performance test record under
"A Return Loss B/R," 3 GHz–6 GHz.
32.Connect the equipment as shown in
Figure 2-20
Receiver R Input Impedance Test Calibration Test Setup
33.Press
A/B
Meas
34.Press
Sweep Setup
Chapter 2
300
k/m
2
M/µ
MARKER 2
1.3
G/n
MARKER 2
3
G/n
Figure
.
TRIGGER MENU
MARKER 2
2
M/µ
MARKER 2
1.3
G/n
to set marker 1 and marker 2
3
G/n
to set marker 1 and marker 2
6
G/n
2-20.
.
CONTINUOUS
Performance Tests
11. Receiver Input Impedance
to set marker 1 and
to set marker 1 and
2-45