Agilent Technologies 8753ES Option 011 Service Manual page 200

Network analyzer
Hide thumbs Also See for 8753ES Option 011:
Table of Contents

Advertisement

Power Supply Troubleshooting
If the Red LED of the A15 Is ON
Verify the A15 Preregulator
Verify that the A15 preregulator is supplying the correct voltages to the A8 post regulator.
Use a voltmeter with a small probe to measure the output voltages of A15W1's plug. Refer
to
Table 5-2
and
Figure
• If the voltages are not within tolerance, replace A15.
• If the voltages are within tolerance, A15 is verified. Continue to
Assembly" on page
Table 5-2
Output Voltages
Pin
A15W1P1 Voltages
(Disconnected)
1
N/C
2
+125 to +100
3, 4
+22.4 to +33.6
−22.4 to −33.6
5, 6
7
N/C
8
+9.4 to +14
−9.4 to −14
9, 10
11
N/C
12
+32 to +48
NOTE
The +5 VD supply must be loaded by one or more assemblies at
all times, or the other voltages will not be correct. It connects to
the motherboard connector A17J3 Pin 4.
5-10
5-6.
5-11.
A8J2 Voltages
(Connected)
+68 to +76.2
+68 to +76.2
+17.0 to +18.4
−17.0 to −18.4
+7.4 to +8.0
+7.4 to +8.0
−6.7 to −7.4
+24.6 to +28.6
+24.6 to +28.6
"Check for a Faulty
A15 Preregulator
Label
N/C
+70 V
+18 V
−18 V
N/C
+8 V
−8 V
N/C
+25 V
Chapter 5

Advertisement

Table of Contents
loading

Table of Contents