Channel B Magnitude Compression
24.Connect the equipment as shown in
Figure 2-28
Channel B Magnitude Compression Test Setup
25.Press
B/R
Meas
26.Repeat the procedure starting with step 4 for the CW frequencies listed in the
performance test record.
In Case of Difficulty
1. If the analyzer fails the channel A compression test, suspect the A5 A sampler
assembly. Repeat this test. Replace the sampler assembly if the problem still exists.
2. Suspect the A6 B sampler assembly if the analyzer fails the channel B compression test.
Repeat this test. Exchange the sampler assembly if the failure still occurs.
Chapter 2
Figure
TRIGGER MENU
Sweep Setup
13. Receiver Magnitude Compression
2-28.
CONTINUOUS
Performance Tests
.
2-61