Node 9
+0.37 V (+0.37 V reference)
Perform step A10, above, and then press
.
9
x1
Check for a flat line at approximately +0.37 V. This is used as the voltage reference in
"Analog Bus Correction Constants (Test 46)" on page
same in high and low resolution; the absolute level is not critical.
Node 10
+2.50 V (+2.50 V reference)
Perform step A10, above, and then press
10
x1
Scale Ref
Check for a flat line at approximately +2.5 V. This voltage is used in
Correction Constants (Test 46)" on page 3-9
low resolution circuitry.
Node 11
Aux Input (rear panel input)
Perform step A10 and then press
This selects the rear panel AUX INPUT to drive the analog bus for voltage and frequency
measurements. It can be used to look at test points within the instrument, using the
analyzer's display as an oscilloscope. Connect the test point of interest to the rear panel
AUX INPUT BNC connector.
This feature can be useful if an oscilloscope is not available. Also, it can be used for testing
voltage-controlled devices by connecting the driving voltage of the device under test to the
AUX IN connector. Look at the driving voltage on one display channel, while displaying the
S-parameter response of the test device on the other display channel.
With
switched ON, you can examine the analyzer's analog bus nodes with
AUX OUT
external equipment (see
considerations, see
"GPIB Service Mnemonic Definitions" on page
Node 12
A10 Gnd (ground reference)
This node is used in
"Analog Bus Correction Constants (Test 46)" on page 3-9
for calibrating the analog bus low and high resolution circuitry.
A11 Phase Lock To observe the A11 analog bus nodes, perform step A11, below. Then
follow the node-specific instructions.
Step A11: Press
Preset
ANALOG BUS ON
Node 13
VCO Tune 2 (not used)
Chapter 10
.
1
x1
Meas
AUX OUT on OFF
ANALOG IN
Meas
Format
MORE
REAL
Service Key Menus and Error Messages
ANALOG IN
Meas
3-9. The voltage level should be the
ANALOG IN
Meas
as a reference for calibrating the analog bus
ANALOG IN
11
in
Table 10-12 on page
Marker
System
.
Service Key Menus
RESOLUTION [HIGH]
RESOLUTION [LOW]
"Analog Bus
.
x1
10-21). For GPIB
10-42.
as a reference
SERVICE MENU
10-29