Messages
Error Messages
-257
File name error
A file name error. This message appears when an error exists in the file name
and hence a command is not executed correctly. This error occurs when you try
to copy to an unsuitable file name, for example.
-256
File name not found
The file name specified is not found and hence the command is not executed
correctly. This error occurs when you try to read a file that does not exist in a
disk or a disk is not correctly inserted into the drive to read or write a file, for
example.
107
File transfer failed
This error occurs when writing data into or reading data from a file
(MMEM:TRAN command) fails.
56
Fixture Simulator not allowed
This error occurs when the fixture simulator is executed (CALC:FSIM:STAT ON)
under the condition of the frequency offset is enabled (SENS:OFFS ON).
The fixture simulator is allowed to use when the frequency offset is not
enabled.
24
Frequency offset enabled
This error occurs when the normal calibration (SENS:CORR:COLL:OPEN etc.) is
performed under the condition of the frequency offset is enabled (SENS:OFFS
ON).
25
Frequency offset not enabled
This error occurs when the scalar-mixer calibration
(SENS:CORR:OFFS:COLL:OPEN etc.) is performed under the condition of the
frequency offset is not enabled (SENS:OFFS OFF).
62
Frequency out of range
This error occurs when the setting of frequency exceeds output frequency
range under the condition of the frequency offset is enabled (SENS:OFFS ON).
G
-105
GET not allowed
A group execution trigger (GET) has been received in the program message
(see 7.7, IEEE488.2).
H
-114
Header suffix out of range
The unit of the header is outside the range. The header is invalid in the unit for
numeric parameters following a SCPI command.