Replacement Procedure

RF Switch and Distributor Module Replacement for 4-Port model (T2-SW and A4)

RF Switch and Distributor Module Replacement for 4-Port model
(T2-SW and A4)

Up to 8.5 GHz Options

Tools Required
Removal Procedure
Refer to
Step 1. Remove the outer cover as described in
Step 2. Remove the following assemblies.
Step 3. Disconnect the semi-rigid cable (item 1).
Step 4. Disconnect the semi-rigid cable (item 2).
Step 5. Disconnect the flat cable (item 3) from the distributor module.
Step 6. Remove a TORX T10 screws (item 4) fastening the guard.
Step 7. Remove a TORX T10 screws (item 5) fastening the guard.
Step 8. Lift the guard from the distributor module (item 6).
Step 9. Remove the three TORX T10 screws (item 7) fastening the plate.
Step 10. Disconnect the three switch cables (item 8) from the switch.
Step 11. Remove the six TORX T10 screws (item 9) fastening the switch.
256
TORX screwdriver, T10, T15, and T20
Flat edge screwdriver
Open-end wrench, 5/8 inch
Open-end torque wrench, 5/16 inch or 8mm (set to 1.08 N-m / 9.5 lb-in)
(for reconnecting SMA connector)
Torque driver set, 5/16 inch (set to 1.08 N-m / 9.5 lb-in) (for reconnecting
SMA connector)
Figure 6-31
for this procedure.
Front Panel Assembly (refer to
Removable Hard Disk Assembly (refer to
Removal (Obsolete)" on page
Disconnect all semi-rigid cables.
"Outer Cover Removal" on page
"Front Panel Removal" on page
"Removable Hard Disk Assembly
205.)
Keysight E5071C ENA Network Analyzers
203.
204.)

Hide quick links: