13. Uncorrected System Performance; Description; Test Equipment - Keysight E5071C ENA Service Manual

Network analyzers
Hide thumbs Also See for E5071C ENA:
Table of Contents

Advertisement

Performance Test

13. UNCORRECTED SYSTEM PERFORMANCE

13. UNCORRECTED SYSTEM PERFORMANCE

Description

This test checks the directivity, source match, load match, transmission
tracking, and reflection tracking which are the key S-parameter measurement
hardware characteristics.
These characteristics are tested using the Keysight 85052D (Over 8.5 GHz
Option
segment sweep points for a specified frequency range. After a full two-port
(Options 230, 235, 240, 245, 260, 265, 460, 465, 280, 285, 2D5, and 2K5) or a
full four-port (Options 430, 435, 440, 445, 480 ,485, 4D5, and 4K5) calibration
is performed for each test port, the test will output the directivity, source
match, reflection tracking, and transmission tracking and load match data to
the report.

Test equipment

Test Equipment (DSG)
Calibration Kit, Type-N
(Calibration Kit N-Type 1)
Type-N Cable
(Cable N:M-N:M 1)
Calibration Kit, 3.5 mm
(Calibration Kit 3.5 mm 1)
3.5 mm Cable
(Cable 3.5 mm: M-3.5 mm: M 1)
3.5 mm (f) to 3.5 mm (f) adapter
(Adapter 3.5 mm: F-3.5 mm: F 1)
56
2
) or Keysight 85032F (Up to 8.5 GHz Option
1
1. The DSG (Device Substitution Group) is listed below the equipment descrip-
tion, if applicable. These devices are required to be mapped in TME to execute
a test. DSGs require either GPIB communication or are important for dynamic
measurement uncertainty calculations. N/A refers to required equipment that
is not mapped as DSGs in TME.
2. Over 8.5 GHz are Options 2D5/2K5/4D5/4K5.
3. Up to 8.5 GHz are Options
230/235/430/435/240/245/440/445/260/265/460/465/280/285/480/485.
3
) Calibration Kit with
Model Number
Over 8.5
GHz
Option
850032F
8120-8862
85052D
X
8121-1834
X
85027-60005
X
Keysight E5071C ENA Network Analyzers
Up to 8.5
GHz
2
3
Option
X
X

Hide quick links: