-223
Too much data
The block-, expression-, or character-string-type program data that has been
received conforms with the standard. But it exceeds the amount that can be
processed under the condition of the memory or conditions specific to
memory-related devices. In this instrument, this error occurs when the number
of characters exceeds 254 in a character-string parameter.
54
Transform, Gate not allowed
This message appears when you turn on the gating/conversion function of the
time domain feature, while frequency offset is enabled, measurement points
are set to two, or sweep type is set to other than linear sweep
Turn off the frequency offset (SENS:OFFS OFF command), set the
measurement points to three or more and the sweep type to linear sweep, and
then turn on the gating/conversion function.
-211
Trigger ignored
This instrument receives and detects a trigger command ("TRIG") or an
external trigger signal. But it is ignored due to the timing condition (This
instrument is not in the wait-for- trigger state, for example). Change the setup
so that a trigger command or an external trigger signal can be sent after the
instrument has entered the wait-for- trigger state.
U
300
Unable to estimate adapter length
Adapter length can not be estimated at zero span. Change the frequency span
other 0.
-113
Undefined header
A command not defined in this instrument, though not illegal in the syntactic
structure, has been received. For example, when a message
":DISP:WIND1:TABL:MEM ON" is sent to a correct program message
":DISP:WIND1:TRAC1:MEM ON," the message sent is received as an undefined
command by this instrument. See the command reference and use correct
commands.
This error occurs also when a port not existing on this model is specified in a
command specifying a port number as an index. Such commands are
CALC:FSIM:SEND:DEEM:PORTn:xxxx, CALC:FSIM:SEND:PMC:PORTn:xxxx,
CALC:FSIM:SEND:ZCON:PORTn:Z0:R, and SENS:CORR:EXT:PORTn:TIME; they
include PORTn as a part.
34
User characterization not found in module
This error occurs when the selected user profile is not detected in the ECal
memory, while reading it from the ECal module, written by the user definition
ECal.
356
Keysight E5071C Service Manual