Imix, Attenuator And Distributor Module Replacement (Imx, Att And A4); 2-Ports Options - Keysight E5071C ENA Service Manual

Network analyzers
Hide thumbs Also See for E5071C ENA:
Table of Contents

Advertisement

Replacement Procedure

Imix, Attenuator and Distributor Module Replacement (IMX, ATT and A4)

Imix, Attenuator and Distributor Module Replacement (IMX, ATT and A4)

1) 2-Ports Options

Tools Required
Removal Procedure
Refer to
Step 1. Remove the outer cover as described in
Step 2. Remove the following assemblies.
Step 3. Disconnect the two semi-rigid cables (item 1).
Step 4. Disconnect the four semi-rigid cables (item 2).
Step 5. Remove three TORX T10 screws (item 3) fastening the Imix module.
Step 6. Lift the Imix module.
Step 7. Remove two TORX T10 screws (item 4) fastening the distributor module to
the plate.
Step 8. Lift the attenuator module.
Step 9. Remove four TORX T10 screws (item 5) fastening the guard.
Step 10. Lift the guard from the attenuator module (item 6).
Step 11. Disconnect the three semi-rigid cables (item 7).
Step 12. Remove the four TORX T10 screws (item 8) fastening the plate.
Step 13. Remove the two TORX T10 screws (item 9) fastening the plate.
Step 14. Lift the plate (item 10).
Step 15. Remove the four TORX T10 screws (item 11) fastening the attenuator module.
Step 16. Lift the distributor module (item 12).
258
TORX screwdriver, T10
Open-end wrench, 5/8 inch
Open-end torque wrench, 5/16 inch or 8mm (set to 1.08 N-m / 9.5 lb-in)
(for reconnecting SMA connector)
Torque driver set, 5/16 inch (set to 1.08 N-m / 9.5 lb-in) (for reconnecting
SMA connector)
Figure 6-32
for this procedure.
Front Panel Assembly (refer to
Removable Hard Disk Assembly (refer to
Removal (Obsolete)" on page 205
Disconnect all semi-rigid cables.
"Outer Cover Removal" on page
"Front Panel Removal" on page
"Removable Hard Disk Assembly
.)
Keysight E5071C ENA Network Analyzers
203.
204.)

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents