Messages
Error Messages
-148
Character data not allowed
A character data element (not violating the standard) has been received at a
position where this instrument does not accept one. Double quotes (") are
omitted where it is necessary to place a parameter in double quotes ("), for
example.
-100
Command error
A comprehensive syntax error has occurred showing that this instrument
cannot detect a more detailed error. This code simply shows that a command
error defined in 11.5.1.1.4, IEEE488.2 has occurred.
60
Continuous switching may damage source attenuator
This error occurs when different power ranges are selected in multiple channel
measurement settings to avoid source attenuator damage.
This error occurs in the case of E5071C Over 8.5 GHz Options
D
-222
Data out of range
A data element (not violating the standard) outside the range defined by this
instrument has been received. This error occurs when an integer-based
command for which the parameter can be rounded exceeds the range of
-65536 to +65536 or when a real-number-based command for which the
parameter can be rounded exceeds the range of -9.9e37 to +9.9e37, for
example.
This error occurs also when a numeric value other than a specified one is
entered into a command in which the "port number" and "CalKit number" are
specified as parameters and hence the parameters are not rounded. Such
commands are, for example, CALC:FSIM:BAL:TOP:BBAL:PPOR,
SENS:CORR:COLL:ACQ:OPEN, SENS:CORR:COLL:ECAL:SOLT3,
SENS:CORR:COLL:CKIT:ORD:LOAD, etc.
-104
Data type error
The parser has recognized a data element that must not exist. Block data has
been sent instead of numeric value data or character string data that had been
expected, for example.
E
32
ECal module not in appropriate RF path
This error occurs when an ECal command, SENS:CORR:COLL:ECAL:SOLTn or
SENS:CORR:OFFS:COLL:ECAL:xxxx, is executed with the port on the ECal
module not connected correctly to the instrument. When the auto-detect
Keysight E5071C Service Manual