Keysight E5071C ENA Service Manual page 325

Network analyzers
Hide thumbs Also See for E5071C ENA:
Table of Contents

Advertisement

System Recovery
Initial Registration of E5071C
Step 6. In the next dialog box, set the date and time for your Windows. Then, click
the Next> button.
Figure B-17
Date and Time Setting dialog box
Step 7. After awhile, the E5071C restarts automatically.
Step 8. E5071C firmware is executed, then the measurement display appears.
If the touch screen does not work after Initial Registration, refer to
"Reconfiguring Touch Panel Settings" on page 329
Reconfiguring Touch Panel Settings
Step 1. In E5071C, right click My Computer.
Step 2. Select Properties. The System Properties dialog box appears.
Step 3. Select Hardware tab.
Step 4. Click Device Manager. The Device Manager dialog box appears.
Step 5. Select Ports (COM & LPT).
Step 6. Check if two Communication Port (COM1) are displayed.
Keysight E5071C Service Manual
-325