Calibration Of The Touch Screen - Keysight E5071C ENA Service Manual

Network analyzers
Hide thumbs Also See for E5071C ENA:
Table of Contents

Advertisement

Replacement Procedure
Removable Hard Disk Drive Replacement (Obsolete)

Calibration of the Touch Screen

When you have replaced the HDD on a E5071C equipped, you have to
calibrate the touch screen. Follow the procedure described below to calibrate
the touch screen.
Step 1. Press
Step 2. Press Service Menu.
Step 3. Press Test Menu.
Step 4. Press Adjust Touch Screen.
The touch screen calibration screen
Figure 6-5
Touch Panel Calibration Screen
Step 5. Touch the x mark on the upper left with your finger. The mark x appears also
on the lower left, upper right, and lower right. Touch the x marks in that order
with your finger.
Touching the four locations described above with your finger automatically
concludes the touch screen calibration.
With no operation on the touch screen calibration screen for a preset time,
it automatically closes and the previous measurement screen reappears.
208
.
(Figure
6-5) appears.
Keysight E5071C ENA Network Analyzers

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents