System Recovery
Initial Registration of E5071C
Figure B-21
Keysight Recovery Image Setup Screen(1/2)
Step 7. A progress bar appears
Image setup. The complete process takes less than 5 minutes to finish.
Figure B-22
Keysight Recovery Image Setup Screen(2/2)
Step 8. A message dialog box appears after the completion of the Recovery Image
Setup procedure. The dialog box disappears automatically after a few
seconds.
Step 9. E5071C firmware is executed, then the measurement display appears.
If the touch screen does not work after Initial Registration, refer to
"Reconfiguring Touch Panel Settings" on page
Reconfiguring Touch Panel Settings
Step 1. In E5071C, right click My Computer.
Step 2. Select Properties. The System Properties dialog box appears.
Step 3. Select Hardware tab.
Step 4. Click Device Manager. The Device Manager dialog box appears.
Step 5. Select Ports (COM & LPT).