Equipment Used For The Power Meter Accuracy Test; Description Of The Test - Keysight E5071C ENA Service Manual

Network analyzers
Hide thumbs Also See for E5071C ENA:
Table of Contents

Advertisement

General Information
Power Meter Accuracy Test

Equipment Used for the Power Meter Accuracy Test

Equipment Type
Range calibrator
Precision digital
voltmeter
Power sensor
cable
Cable, BNC, 50 
Adapter for
connecting BNC
cable to DVM
inputs

Description of the Test

The power meter accuracy is verified for various power inputs and the actual
readings are recorded in a test record. A range calibrator is used to provide the
reference inputs.
It is recommended that a copy of the test record on
the values be recorded on the copy, thus preserving the original for future
use.
Test Procedure
This procedure assumes the use of the recommended equipment model
numbers. The actual steps required, therefore, may differ for other model
numbers of equipment used.
Step 1. Setup the equipment as shown in
a. Connect the DC REFERENCE OUTPUT connector on the rear panel of the
b. Connect the POWER METER output of the range calibrator to the input of
c. Switch on the power to the power meter, the range calibrator, and the
32
Recommended Model
or Part Number
Keysight 11683A
Keysight 3458A
Keysight 11730A, p/n
8120-8319
Any
Any
1. Required accuracy and resolution at the following voltage levels:
14mV input: 0.0100% accuracy, 10nV resolution
0.14V input: 0.0050% accuracy, 100nV resolution
0.45V input: 0.0030% accuracy, 100nV resolution
range calibrator to the DVM voltage input.
the power meter being tested.
digital voltmeter.
Alternate
Model or Part Number
None
Any with the required
accuracy and resolution
Any equivalent
Any
Any
page 36
Figure
1-1.
Keysight E5071C ENA Network Analyzers
1
be made, and