Messages
Error Messages
S
106
Save failed
This error occurs when writing an instrument status file (State01.sta, etc.)
(MMEM:STOR:STAT command) fails.
33
Selected parameter not valid for confidence check
This error occurs when the mix mode S parameter has been selected for the S
parameter you want to check, while using the confidence check function for
calibration coefficient.
57
SnP request not valid for selected measurement
This error occurs when you try to save data to a Touchstone file but no
measurement has been executed.
This error also occurs when you try to save a Touchstone file with power sweep
measurement specified or with the frequency offset function set to ON.
50
Specified channel hidden
This error occurs when an attempt is made to activate a channel not on display
using the DISP:WIND:ACT command. This error is not generated by front key
operations.
23
Specified error term does not exist
The error occurs when the calibration coefficient type of the parameter
specified in the read/write command of the calibration coefficient
(SENS:CORR:COEF) is invalid for the calibration type selected by the
calibration type selection command (SENS:CORR:COEF:METH:xxxx). This error
is not generated by front key operations.
21
Specified ports overlapped
This error occurs when a port number is duplicated in a command requiring
two or more port numbers as parameters. Such commands are, for example,
CALC:FSIM:BAL:TOP:SSB:PPOR 1,2,3,3. Specify port setup correctly to avoid
duplication of ports. This error is not generated by front key operations.
51
Specified trace dose not exist
This error occurs when CALC:PAR:SEL command is executed to activate more
traces than specified by CALC:PAR:COUN command.This error is not
generated by front key operations.
26
Standard not selected
This error occurs when you execute the measurement under the condition that
the standard is not selected for the subclass (the standard number is set to 0).
-150
String data error
When a character-string-data element is put to syntactic analysis, an error not
corresponding to one of Error Numbers -151 through -159 occurs.