Keysight E5071C ENA Service Manual page 35

Network analyzers
Hide thumbs Also See for E5071C ENA:
Table of Contents

Advertisement

General Information
Power Meter Accuracy Test
Step 19. Record the power meter reading as value E in the test record.
Step 20. If testing a dual-channel power meter, perform
the other channel.
Step 21. Perform the pass/fail calculations indicated on the test record.
If a channel of the power meter does not pass this test, the power meter
cannot be used in applications that require Option G12 or H12. There are
no adjustments that can be performed to improve the performance of the
power meter. Typically, replacing the A86 measurement assembly
associated with the failed channel will correct the problem.
Keysight E5071C ENA Network Analyzers
Step 7
through
Step 19
for