Keysight E5071C ENA Service Manual page 345

Network analyzers
Hide thumbs Also See for E5071C ENA:
Table of Contents

Advertisement

Messages
Error Messages
104
Failed to create directory
This error occurs when creating a directory (MMEM:MDIR command) fails.
103
Failed to delete file
This error occurs when deleting a file (MMEM:DEL command) fails.
100
Failed to read file
This error occurs when a 2-port touchstone file
(CALC:FSIM:SEND:PMC:PORT:USER:FIL command), the formatted data array
(MMEM:LOAD:FDAT command) and limit table (MMEM:STOR:LIM command)
for the active trace on the active channel, segment sweep table
(MMEM:LOAD:SEGM command) for the active channel, a VBA project file
(MMEM:LOAD:PROG command), etc. cannot be read normally.
101
Failed to write file
This error occurs when the formatted data array (MMEM:STOR:FDATcommand)
and limit table (MMEM:STOR:LIM command) for the active trace on the active
channel, segment sweep table (MMEM:STOR:SEGM command) for the active
channel, display image (MMEM:STOR:IMAG command) for the LCD screen, a
VBA project file (MMEM:STOR:PROG command), etc. cannot be written
normally.