General Information
Power Meter Accuracy Test
Test Record for Power Meter Accuracy Test
Power Meter Tested
Model Number:
Serial Number:
Test Equipment Used
Range Calibrator: Model No.:
Digital Voltmeter: Model No.:
Test Results
Range
Calibrator:
Setting
1 mW
300 W
100 W
Pass/Fail Calculation
300 W
100 W
36
Channel A
DVM Reading
Power Meter
(Vdc)
Reading (dBm)
A=
0.000.001
B =
C =
D =
E=
R=B/A =
(C/10)
S=10
=
% ERROR=((R-S)/R)x100 =
Limits: 0.13%
Pass __ Fail __
T=D/A =
(E/10)
U=10
=
% ERROR=((T-U)/T)x100 =
Limits: 0.10%
Pass __ Fail __
Test Date:
Tested by:
Serial No.:
Serial No.:
Channel B
DVM Reading
Power Meter
(Vdc)
Reading (dBm)
A=
0.00 0.001
B =
C=
D =
E =
R=B/A =
(C/10)
S=10
=
% ERROR=((R-S)/R)x100 =
Limits: 0.13%
Pass __ Fail __
T=D/A =
(E/10)
U=10
=
% ERROR=((T-U)/T)x100 =
Limits: 0.10%
Pass __ Fail __
Keysight E5071C ENA Network Analyzers