Replacement Procedure

USB (USBTMC) Interface Card Replacement (Serial number prefix MY461 or SG461)

USB (USBTMC) Interface Card Replacement (Serial number prefix
MY461 or SG461)

Tools Required

Removal Procedure

Refer to
Step 1. Remove the outer cover as described in
Step 2. Remove the three TORX T10 screw (item 1) fastening the angle and lift it.
Step 3. Disconnect the flat cable (item 2) from the PCI DSP card.
Step 4. Remove the two TORX T10 screws (item 3) fastening the card to the rear
panel.
Step 5. Disconnect the assembly (item 4) from the Digital Motherboard and lift it.
Step 6. Remove the two TORX T10 screws (item 5) fastening the angle.
224
TORX screwdriver, T10, T15, and T20
Figure 6-16
for this procedure.
"Outer Cover Removal" on page
Keysight E5071C ENA Network Analyzers
203.