Warning Message - Keysight E5071C ENA Service Manual

Network analyzers
Hide thumbs Also See for E5071C ENA:
Table of Contents

Advertisement

Messages

Warning Message

Warning Message
A warning message is displayed in the instrument message/Warning area in
the lower left part of the display against a gray background. Pushing a front
panel key or executing :DISP:CCL command clears the message.
This message simply appears on the display, being not known to a remote
environment such as a GPIB. This message is not displayed when another error
(against a red background) has already been displayed in the instrument
message/Warning area.
The warning messages for this instrument are as follows:
Auto port extension not allowed
This message appears when you try to execute the auto port extension when
the sweep type is the power sweep or the frequency offset function is ON.
If this message is displayed, the auto port extension is not available.
Calibration extrapolated
The frequency range that can be calibrated with the ECal module is exceeded.
In this case, the calibration data for the minimum frequency or maximum
frequency are used instead of the data exceeding the allowable frequency
range for calibration.
Fixture simulator not allowed
This message appears when enabling the frequency offset, while the fixture
simulator function is turned on.
If this message is displayed, the fixture simulator function is automatically
turned off.
Log sweep requires 2 octave minimum span
If you change the sweep range that does not satisfy the necessary condition of
the frequency span (the stop frequency is about 4 times or more of the start
frequency) when the sweep type is set to the log sweep, this message is
displayed.
If this message is displayed, the sweep type is automatically set to the linear
sweep.
For example, if you attempt to change the log type to the log sweep when the
start frequency is set to 1 MHz and the stop frequency to 2 MHz, this error
occurs.
Set the stop frequency to a value of about 4 times or more the start frequency
and then set the low seep.
Peak not found
This warning message is displayed when, with the tracking turned on, the peak
specified by the marker search has not been found by the time the sweep is
finished (with the tracking executed).
Keysight E5071C Service Manual
358