Replacement Procedure
RF Switch and Distributor Module Replacement for 2-Port model (A4)
Figure 6-30
RF Switch and Distributor Module Replacement for 2-port model
Replacement Procedure
Step 1. Replace the switch and distributor module by inverse procedure of removal.
Be careful screw (item 2) and screw (item 3) are different.
Fasten the semi rigid cables, attenuator and terminations using a torque
wrench or torque driver set for replacement.