Replacement Procedure

Outer Cover Removal

Outer Cover Removal

Tools Required

Procedure

Refer to
Step 1. Disconnect the power cable from the E5071C.
Step 2. Remove the four bottom feet (item 1).
Step 3. Remove the four TORX T10 screws (item 2) from the bottom.
Step 4. Remove the four TORX T20 screws (item 3) fastening the side strap handles.
Step 5. Remove the four TORX T15 screws (item 4) fastening the four rear foot
Step 6. Remove the five TORX T15 screws (item 5) fastening the cover.
Step 7. Slide up the outer cover and remove it carefully.