Pci Dsp Card And Gpib Card Replacement (A51) (Serial Number Prefix My461 Or Sg461).226; Tools Required; Removal Procedure - Keysight E5071C ENA Service Manual

Network analyzers
Hide thumbs Also See for E5071C ENA:
Table of Contents

Advertisement

Replacement Procedure
PCI DSP Card and GPIB Card Replacement (A51) (Serial number prefix MY461 or
SG461)
PCI DSP Card and GPIB Card Replacement (A51) (Serial number prefix
MY461 or SG461)

Tools Required

Removal Procedure

Refer to
Step 1. Remove the outer cover as described in
Step 2. Remove the three TORX T10 screw (item 1) fastening the angle and lift it.
Step 3. Remove the two TORX T10 screw (item 2) fastening the stay and lift it.
Step 4. Disconnect the flat cables (item 3).
Step 5. Disconnect the flat cable (item 4) from the handler I/O board.
Step 6. Remove the four TORX T10 screw (item 5) fastening the card to the rear
panel.
Step 7. Disconnect the USB(USBTMC) interface card (item 6) from the Digital
Motherboard and lift it.
Step 8. Disconnect the assembly (item 7) from the Digital Motherboard and lift it.
Step 9. Remove the PCI DSP Card or GPIB Card.
For removing the PCI DSP Card.
For removing the GPIB Card.
226
TORX screwdriver, T10, T15, and T20
Open-end torque wrench, 5/8 inch (set to 1.97 N-m / 17.4 lb-in)
Figure 6-17
for this procedure.
Remove the two 5/8 inch nuts (item 8) fastening the angle.
Remove a TORX T10 screw (item 9) fastening the angle.
Remove the two TORX T10 screws (item 10) fastening the angle.
"Outer Cover Removal" on page
Keysight E5071C ENA Network Analyzers