Keysight E5071C ENA Service Manual page 359

Network analyzers
Hide thumbs Also See for E5071C ENA:
Table of Contents

Advertisement

Messages
Warning Message
Segment table changed
This warning message is displayed when the setting specified segment by
segment in the segment table is automatically changed by a change in the
other setting.
For example, this warning message is displayed when, with the power
specified segment by segment in the segment table, the power setting for a
segment is adjusted by a change in the power range setting.
Sweep mode changed to stepped sweep
This message appears when you set the sweep type to power sweep, power
slope function to enabled (ON, correction coefficient is other than zero), power
calibration to ON, frequency offset to enabled, spurious avoidance of frequency
offset to enabled, or external signal source control to enabled, while the sweep
mode is set to swept mode or high speed swept mode.
When this message is displayed, the sweep mode is automatically set to step
mode (during swept mode) or high speed step mode (during high speed swept
mode).
Target value not found
This warning message is displayed when, with the tracking turned on, the
target specified by the marker search has not been found by the time the
sweep is finished (with the tracking executed).
This warning message is displayed also when, with the bandwidth marker
displayed, the setting for the bandwidth marker is changed at the end of the
sweep, or when, with the active marker changed or moved, the bandwidth is
not found.
Transform, Gate not allowed
This message appears when you have set the measurement points to two,
sweep type to other than linear sweep, or frequency offset to enabled, while
the gating/conversion function of the time domain feature is turned on.
If this message is displayed, the gating/conversion function is automatically
turned off.
User Preset File not found
This warning message is displayed if no user preset file (D:\UserPreset.sta)
exists when executing the SYST:UPR command.
If this message is displayed, a normal preset (SYST:PRES) will be performed.
Keysight E5071C Service Manual
-359

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents