Post-Repair Procedures
Post-Repair Procedures
Table 7-2
Post-Repair Procedures for Analog Assemblies for Over 8.5 GHz
Options
Replaced
Assembly or
Part
Synthesizer
Module (A1)
Receiver
Module (A7)
Front End
Module (A9)
Keysight E5071C ENA Network Analyzers
Required Adjustments
Correction Constants
(CC)
Perform the following
required adjustment.
- Synthesizer Gain
- Source Output Power
- Receiver Port
Characteristics
- Receiver Absolute Gain
Perform the following
required adjustments.
- Source Output Power
- Receiver IF Range
- Receiver Port
Characteristics
- Receiver Absolute Gain
Perform the following
required adjustments
- Source Output Power
- Receiver Port
Characteristics
- Receiver Absolute Gain
Verification
Perform
"To Execute the diagnostics
Test" on page 80
Perform the following performance test.
- Frequency Accuracy Test
- RF Output Level Accuracy and Flatness
Test
- RF Output Level Linearity Test
- Trace Noise Test
- Crosstalk Test
- System Dynamic Range Test
- Noise Floor Test
- Dynamic Accuracy test
- Uncorrected System Performance Test
Perform
"To Execute the diagnostics
Test" on page 80
Perform the following performance test.
- Trace Noise Test
- Crosstalk Test
- System Dynamic Range Test
- Noise Floor Test
- Dynamic accuracy test
- Uncorrected system performance
Perform
"To Execute the diagnostics
Test" on page 80
Perform the following performance test.
- Uncorrected System Performance Test
289