How to exit from the E5071C Measurement View

How to exit from the E5071C Measurement View
You need to exit from the E5071C Measurement View to perform some
troubleshooting. The following is the procedure to exit from the E5071C
Measurement View.
Step 1. Connect the mouse and external keyboard to the connectors on the E5071C
rear panel.
Step 2. Turn the instrument on.
Step 3. Press
Step 4. Click Service Menu - Service Function. Password dialog box opens as shown
in
Figure 4-1
Password dialog box
Step 5. Enter the password kid in the Password box and then click "OK" button.
Step 6. Click Exit in Service Functions Menu. Then the E5071C exit the Measurement
View, then windows desktop screen appears with 3 icons (Network Analyzer,
Recycle Bin, and Internet Explorer.)
If you wish to return to Measurement View, double-click "Network
Analyzer" icon.
If you need to shut down the E5071C and again turn on, perform in
accordance with the following procedure.
a. To get "Start" menu bar displayed, move the pointer to the bottom of the
b. Click "Start" and "Turn off Computer" in the pull down menu. "Turn off
c. Click "Turn off" button in the dialog box.
66
key.
Figure
4-1.
screen with mouse.
Computer" dialog box opens.