Synthesizer Module Replacement (A1); Tools Required; Removal Procedure - Keysight E5071C ENA Service Manual

Network analyzers
Hide thumbs Also See for E5071C ENA:
Table of Contents

Advertisement

Replacement Procedure

Synthesizer Module Replacement (A1)

Synthesizer Module Replacement (A1)

Tools Required

Removal Procedure

Refer to
Step 1. Remove the outer cover as described in
Step 2. Remove the front panel as described in
Step 3. Disconnect the semi-rigid cable (item 1).
Step 4. Disconnect the semi-rigid cable (item 2).
Step 5. Remove the two TORX T10 screws fastening the module to the bottom. (item
3)
Step 6. Slide the synthesizer module (item 4).
240
TORX screwdriver, T10, T15, and T20
Flat edge screwdriver
Open-end wrench, 5/8 inch
Open-end torque wrench, 5/16 inch or 8mm (set to 1.08 N-m / 9.5 lb-in)
(for reconnecting SMA connector)
Torque driver set, 5/16 inch (set to 1.08 N-m / 9.5 lb-in) (for reconnecting
SMA connector)
Figure 6-25
for this procedure.
"Outer Cover Removal" on page
"Front Panel Removal" on page
Keysight E5071C ENA Network Analyzers
203.
204.